Share:


Standard Oil

BIUGED supply a series of different viscosity standard oils which can bring convenience for calibrating your viscometers or flow cups. They are Newton liquid, with a perfect stability, hydrophobic and damp-proof. Good physical inertia and low surface tension. They can be kept and used fora long time.
 
Every bottle of standard oil has been calibrated by SOUTH CHINA NATIONAL CENTER OF METROLOGY, and comes with a calibration certificate report which calibrates dynamic viscosity
(cP valueand kinematic viscositycSt value, and has a very high reliability and accuracy.

 

     Ordering
   Information

    Dynamic Viscosity
              
cP

Kinematic Viscosity

mm2/s

Suitable for

BGD 1600/1

5

5.4

ISO Flow Cup No.3

BGD 1600/2

50

52.1

Zahn Cup No.1Ford Cup No.2

BGD 1600/3

100

104.2

Zahn Cup No.2ISO Flow Cup No.4

Ford Cup No.3Iwata Cup No.2NK-2

BGD 1600/4

200

208.3

ISO Flow Cup No.5

DIN Flow Cup No.4Ford Cup No.4

BGD 1600/5

50067.2 KU

520.8

Zahn Cup No.3 and No.4ISO Flow Cup No.6

BGD 1600/6

1,00086.3 KU

1,052.3

Zahn Cup No.5Krebs Stormer Viscometer

BGD 1600/7

5,000

5,170.6

Rotary Viscometer

BGD 1600/8

10,000

10,416.7

Rotary Viscometer

BGD 1600/9

30,000

31,039.8

Rotary Viscometer

BGD 1600/10

60,000

62,176.2

Rotary Viscometer

BGD 1600/11

100000

103,092.8

Rotary Viscometer whose range is over 100,000 cp

BGD 1600/L

67.2 KU

----

Krebs Stormer Viscometer

BGD 1600/M

86.3 KU

----

Krebs Stormer Viscometer

BGD 1600/H

106.6 KU

----

Krebs Stormer Viscometer

BGD 1600/T

121.5 KU

----

Krebs Stormer Viscometer


Notice:
 ① The tested value of below table is tested at 25℃

② For every bottle standard oil, the fact viscosity value should be taken as final by calibration certificate report.
③ The period of validity for viscosity value of standard oil is one year, it should be recalibrated every year.



 Inquiry - Standard Oil