1

Share:


SO2 H2S Gas Corrosion Test Chamber (SO-225)

SO2 H2S Gas Corrosion Test Chamber (SO-225)

SO2 Noxious Gas Test Chambers reproduce damages by temperature, relative humidity and noxious gas Corrosion to material, components and constructions.

This noxious gas chamber can be used for single gas tests and mixed gas tests.Test methods are in accordance with IEC 60068-2-42, IEC60068-2-60 and other standards.
 

Model SO-100 SO-225 SO-500 SO-800 SO-1000
Internal Dimensions (mm) 400*500*500 500*600*750 700*800*900 800*1000*1000 1000*1000*1000
Overall Dimensions (mm) 860*1050*1620 960*1150*1860 1180*1350*2010 1280*1550*2110 1500*1550*2110
Interior Volume (L) 100 225 500 800 1000
Parameter Temperature Range 15 ℃ ~ 80 ℃
Temperature Fluctuation ± 0.5 ℃
Temperature Deviation ± 2.0 ℃
 Humidity Range 30% ~ 100% RH
Humidity Deviation ±1% RH
SO2 Concentration 0.067% (0.15Liters)
Air Change Rate 3~5 /h
Structure Cooling Cooling system Mechanical compression refrigeration system
Refrigerating unit French TECUMSEH compressor
Refrigerant R404A
Heating Element Nichrome heater
Humidity Humidifier External isolation, stainless steel surface evaporation humidifier
Water Supply System Water purification system
Automatic water supply
Dehumidification Evaporator
Controller Programmable color LCD touch screen controller
Ethernet connection, PC Link
 Sensor Temperature Sensor PT100Ω / MV A-class
 Humidity Sensor Dry and wet bulb sensor
Gas Sensor SO2 gas sensor
Build-in Water Tank(mm) 270*300*400
View Window Size(mm) 355*435
Air Circulation Centrifugal wind fan
Door Lock Electromagnetic lock
Gas Exhaust Device SO2 Exhaust Unit
Safety Device Humidifier Dry-combustion Protection; Over-temperature Protection; Over-current Protection;
Water Shortage Protection; Earth leakage Protection
Interior material SUS304 stainless steel +Glass fiber reinforced plastics
Exterior material Steel Plate with protective coating
Power Supply 380V 60Hz 3 phase
Maximum Noise        65 dBA
Standard ISO 22479 method B
Environmental Conditional 5℃~+35 ℃   ≤85% RH



 Inquiry - SO2 H2S Gas Corrosion Test Chamber (SO-225)